I also have been having low temp errors. I use Benchmark Ultras and they are
not a batch stainer, so pads are heated independently. I noticed the staining
on the particular pad with the error was not quite to
ubject: RE: [Histonet] Low Temp question
We see these continually as well. We are told that it is a software glitch and
not really a problem with the heat. I was told that they are working on it.
Tom McNemar, HT(ASCP)
Histology Co-ordinator
Licking Memorial Health Systems
(740) 348-4163
(740) 348-
www.LMHealth.org
-Original Message-
From: histonet-boun...@lists.utsouthwestern.edu
[mailto:histonet-boun...@lists.utsouthwestern.edu] On Behalf Of Shawn Leslie
Sent: Tuesday, October 22, 2013 11:07 AM
To: Histonet@lists.utsouthwestern.edu
Subject: [Histonet] Low Temp question
Goodmorning all
Goodmorning all,
I have a question concerning the Ventana Benchmark XT. We are constantly
having low temperature errors during almost every run. The immunos look fine
but we are still concerned. Has anyone else that uses the Benchmark experienced
this also ? We have noticed that the silver