RE: [Histonet] Low Temp question

2013-10-22 Thread John Shelley
;Shawn Leslie '; Histonet@lists.utsouthwestern.edu Subject: RE: [Histonet] Low Temp question I also have been having low temp errors. I use Benchmark Ultras and they are not a batch stainer, so pads are heated independently. I noticed the staining on the particular pad with the error was not quite to

RE: [Histonet] Low Temp question

2013-10-22 Thread HERRINGTON, SHEILA
ubject: RE: [Histonet] Low Temp question We see these continually as well. We are told that it is a software glitch and not really a problem with the heat. I was told that they are working on it. Tom McNemar, HT(ASCP) Histology Co-ordinator Licking Memorial Health Systems (740) 348-4163 (740) 348-

RE: [Histonet] Low Temp question

2013-10-22 Thread Tom McNemar
www.LMHealth.org -Original Message- From: histonet-boun...@lists.utsouthwestern.edu [mailto:histonet-boun...@lists.utsouthwestern.edu] On Behalf Of Shawn Leslie Sent: Tuesday, October 22, 2013 11:07 AM To: Histonet@lists.utsouthwestern.edu Subject: [Histonet] Low Temp question Goodmorning all

[Histonet] Low Temp question

2013-10-22 Thread Shawn Leslie
Goodmorning all, I have a question concerning the Ventana Benchmark XT. We are constantly having low temperature errors during almost every run. The immunos look fine but we are still concerned. Has anyone else that uses the Benchmark experienced this also ? We have noticed that the silver