Hi Adrian, 
               We use the Fluke IDA4 here, we have 3 of them in our Dept. it 
comes with 2 or four channel options. For most of our infusion devices this is 
fine, although with devices such as ALARIS P5000 we have to run the occlusion 
test using a force gauge. which is measured in kg/f as opposed to mmHg.
You can also hook a printer up to the IDA4, which is good to traceability/QC. 
 
Best Regards
 
Dom 


Date: Fri, 27 Jun 2008 09:59:21 +0930From: [EMAIL PROTECTED]: [EMAIL 
PROTECTED]: [bmelist] Infusion device analysers

Hi I am wondering what people may have been purchasing recently in terms of 
infusion analysers. We have a home grown one that is currently in its 
twighlight hours and need to go shopping for a replacement.
We need it to cover the broadest range of flow rates, be able to handle 
multiple channels simultaneously and have a computer/management system 
interface of some description
Thanks Adrian Richards 
_________________________________________________________________

http://clk.atdmt.com/UKM/go/msnnkmgl0010000009ukm/direct/01/