Dear Peter,

This noise could be due to the presence of some air in the pump. Before call 
any service, try to flux about 100 ml of 100% methanol (or ethanol if you don’t 
like to use methanol) at 10 ml/min and 5 ml/min (50:50).
This should be remove any bubble. If the noise persist, the pump probably need 
a cleaning and maintenance.

Best
Roberto

___________________________________
Roberto  Orru, PhD
Biochemistry Department, Emory University
1510 Clifton Road
Rollins Research Center Room 4078
Atlanta, GA. 30322 (USA)
Phone: 404.727.5971
Fax: 404.727.2738


-----Original Message-----
From: CCP4 bulletin board [mailto:CCP4BB@JISCMAIL.AC.UK] On Behalf Of Peter Hsu
Sent: Thursday, July 12, 2012 5:09 PM
To: CCP4BB@JISCMAIL.AC.UK
Subject: [ccp4bb] offtopic: AKTA prime

Hi all,

 We've got an old Akta prime that I think is on the verge of kicking it. 
Hearing some high pitched sounds coming from the pump when we're running it. 
Line A seems clogged and makes a thudding noise when we try to do a pump wash 
through that line. Does anyone have any experience w/fixing one/replacing the 
pump? Or any idea how much it's going to set us back to get it fixed?

Sorry for the off topic question/thanks in advance for any thoughts and ideas.

Peter

________________________________

This e-mail message (including any attachments) is for the sole use of
the intended recipient(s) and may contain confidential and privileged
information. If the reader of this message is not the intended
recipient, you are hereby notified that any dissemination, distribution
or copying of this message (including any attachments) is strictly
prohibited.

If you have received this message in error, please contact
the sender by reply e-mail message and destroy all copies of the
original message (including attachments).

Reply via email to