As far as I know it could happen when you are using i2. We are working to fix 
this problem. At the moment the best solution is to use advanced options in the 
refmac interface and define resolution limits explicitly. For example you can 
add in the advanced options:


resolution 37 2.5

where these resolution limits are from the aimless output. Low resolution limit 
should not affect refinement behaviour, however it may cause problem during 
deposition.

We will have better solution soon. I2 has many good features. Please report if 
you see some misbehaviours.

Best wishes,
Garib

P.S. you can also use advanced options to do things that are not available on 
the interface yet.
1) refinement against electron diffraction. Add

source EC MB   #  electron form factor will be used using Mott-Bethe formula

2) you can add occupancy refinement

etc