Hi Ben,

Yes, the CMRR multiband sequence incorporates "dynamic off-resonance in 
k-space” (DORK)” method as in Pfeuffer et al., MRM 2002, which has already been 
implemented in the Siemens product EPI sequence ICE recon.

Gordon
 
> On Jun 8, 2017, at 12:05 PM, Benjamin Risk <bb...@cornell.edu> wrote:
> 
> Dear HCP experts,
> 
> I am trying to better understand the unprocessed fMRI data. Does the online 
> reconstruction include corrections for spatial shifts due to B0 fluctuations? 
> There is a note about this in Ugurbil et al 2013 but I am wondering if this 
> applies to the final HCP acquisition protocol. 
> 
> Thank you!
> Ben Risk
> 
> _______________________________________________
> HCP-Users mailing list
> HCP-Users@humanconnectome.org
> http://lists.humanconnectome.org/mailman/listinfo/hcp-users
> 


_______________________________________________
HCP-Users mailing list
HCP-Users@humanconnectome.org
http://lists.humanconnectome.org/mailman/listinfo/hcp-users

Reply via email to