Error code 73: Symptom: Rotary valve does not stop at the required station after 2 times rotation Probable cause: 1. Photointerrupter defective 2. Positioning disc broken 3. Photointerrupter for position detection disturbed by strong light source Remedy: 1. Replace photointerrupter PC board 2. Replace positioning disc 3. Eliminate light source
Just found this information in my older VIP E300 manual. Lynne A. Bell, HT (ASCP) Technical Specialist, Histology Central Vermont Medical Center 130 Fisher Road Barre, VT 05641 802-371-4923 _______________________________________________ Histonet mailing list Histonet@lists.utsouthwestern.edu http://lists.utsouthwestern.edu/mailman/listinfo/histonet