I am running into troubles with R1rho dispersion fits. In brief, we get totally 
unrealistic values for phi_ex (>10000) and k_ex (<1 s-1).
It seems that the default parameter limits for phi_ex are such that there is 
only a lower bound (0) but no upper bound. This can result in physically 
totally meaningless situations, like a strong R1rho dispersion profile, with a 
fitted exchange rate constants of 0.2 s-1….
I would like to have some control over the grid search and the analysis.

My questions:
* does anyone know how to set/change the grid parameter range? Has anyone run 
into similar issues?
* does anyone have a script to do the analysis without going through the 
auto-analysis? The manual is (for me) not useful to design the analysis.

Thanks

Paul