Palo Verde provides the following response: Palo Verde uses the Bicron Micro Rem Meter and we calibrate on the X1 position using a Shepherd Model 89 Calibrator and expose the detector to a 150 μR/hr gamma field.
Seth. From: [email protected] [mailto:[email protected]] On Behalf Of Pell, Bryan Sent: Thursday, February 03, 2011 8:38 AM To: '[email protected]' Subject: Powernet: MicroRem Calibration Just a quick question. Ginna calibrates our microrem meters on the x1 scale electronically instead of calibrating on that scale against a source. We calibrate the other scales (x10, x100, x1000) against a source. This method is authorized in the tech manual but was challenged. What does your site do? Thanks, Bryan Bryan S. Pell R.E. Ginna Nuclear Power Plant Senior Plant Health Physicist (585) 771-3804 [email protected] P Please consider the environment before printing >>> This e-mail and any attachments are confidential, may contain legal, >>> professional or other privileged information, and are intended solely for >>> the addressee. If you are not the intended recipient, do not use the >>> information in this e-mail in any way, delete this e-mail and notify the >>> sender. CEG-IP1 --- NOTICE --- This message is for the designated recipient only and may contain confidential, privileged or proprietary information. If you have received it in error, please notify the sender immediately and delete the original and any copy or printout. Unintended recipients are prohibited from making any other use of this e-mail. Although we have taken reasonable precautions to ensure no viruses are present in this e-mail, we accept no liability for any loss or damage arising from the use of this e-mail or attachments, or for any delay or errors or omissions in the contents which result from e-mail transmission.