Hi Edward.

Thanks for the comment!

According to manual "In all other cases (P1 is not zero),this scale
factor should be 1.0".

So for P1 = -21.0 I should set -c 1.0 ?

Best
Troels

Troels Emtekær Linnet


2013/8/27 Edward d'Auvergne <[email protected]>:
> Hi,
>
> I've noticed that the NMRPipe processing script has an error in it.
> If you have p1 phasing not equal to zero, then the first point should
> not be halved with '-c 0.5'.  See the relaxation curve-fitting part of
> the relax user manual or
> http://www.nmr-relax.com/manual/Spectral_processing.html for more
> details on the first point scaling.
>
> Regards,
>
> Edward
>
>
>
> On 27 August 2013 11:25, Troels Emtekær Linnet <[email protected]> wrote:
>> Hi.
>>
>> I made a wiki page for at Tutorial.
>> Relaxation dispersion analysis cpmg fixed time recorded on varian as
>> fid interleaved
>>
>> http://wiki.nmr-relax.com/Tutorial_for_Relaxation_dispersion_analysis_cpmg_fixed_time_recorded_on_varian_as_fid_interleaved
>>
>> It is targeted new students in our lab, who can have a little trouble
>> handling linux.
>> It therefore contains some helper scripts, to help processing spectra.
>>
>> Any comments would be appreciated. :-)
>>
>> Troels Emtekær Linnet
>>
>> _______________________________________________
>> relax (http://www.nmr-relax.com)
>>
>> This is the relax-devel mailing list
>> [email protected]
>>
>> To unsubscribe from this list, get a password
>> reminder, or change your subscription options,
>> visit the list information page at
>> https://mail.gna.org/listinfo/relax-devel

_______________________________________________
relax (http://www.nmr-relax.com)

This is the relax-devel mailing list
[email protected]

To unsubscribe from this list, get a password
reminder, or change your subscription options,
visit the list information page at
https://mail.gna.org/listinfo/relax-devel

Reply via email to