Hold that thought.
On closer examination of the difference plot at high angles you have something 
that looks like sample displacement, and refining it improves the fit, although 
the values for zero point and shift become quite large.  Mirrors shouldn't give 
you this kind of error so make sure it's properly aligned and if it's an old 
retrofitted theta-theta system, make sure that your goniometer worm gears 
aren't getting slack.