Hi all,
thanks a lot for all the responses!
As anticipated, the solution is rather simple - the origin of the beam:
denzo refines to: x=82.6 y=80.3
mosflm reads from header: x=80.0 y=84.0 - bad indexing etc
mosflm works with: x=82.6 y=80.3
No swapping of x and y between denzo and mosflm in this cas
Hi
I'd guess from the input file that the beam centre may be defined in a
different reference frame to that which Mosflm expects in the image
header; other possibilities are that the distance, wavelength, or
something else (!) is wrong. Indexing is critically dependent on having
the beam cent
Is it possible that your crystal is not really a single crystal, or has
multiple domains? I had a similar problem with an otherwise beautifully
diffracting crystal (beyond 2.0 A) whose integration rapidly broke down
after a few frames. On close inspection, it was possible to visualize
multiple
Hi all,
processing data collected on APS beamline BM17 on a MAR165 detector
causes unexpected troubles. Even though the crystals are
well-diffracting and the spots are sharp and well-resolved, indexing
only works with a certain selection of frames, the refinement the goes
totally hairwire. Probalby