Re: [Histonet] Hard Water and Auto Stainers

2017-12-20 Thread Morken, Timothy via Histonet
Sandra, yes, we use DI water from a water system in our lab

Tim Morken
Pathology Site Manager, Parnassus 
Supervisor, Electron Microscopy/Neuromuscular Special Studies
Department of Pathology
UC San Francisco Medical Center


-Original Message-
From: Sandra Cheasty via Histonet [mailto:histonet@lists.utsouthwestern.edu] 
Sent: Wednesday, December 20, 2017 10:44 AM
To: Histonet (histonet@lists.utsouthwestern.edu)
Subject: [Histonet] Hard Water and Auto Stainers

Hello all,
Thanks for everyone's response to error 90 on our Prisma 
stainer; it has been cleared! Does anyone use filtered, softened, or DI water 
in their H stainer to avoid clogging up the system?
Thanks!
Sandy

Sandra J. Cheasty, HT (ASCP)
Histology & Necropsy Supervisor
UW-Madison, School of Veterinary Medicine

___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet


___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet


[Histonet] FW: Recycling formalin

2017-12-20 Thread Morken, Timothy via Histonet

Martha, The question to ask your EHS is how much extra exposure they will allow 
employees. Recycling removes all the salts so you need to add them back. Then 
it is just like making up formalin from scratch. Does anyone do that anymore? 
You need a minimum of full-face vapor masks to prevent vapor exposure. And risk 
of spill is high. Our EH  decided it was not worth the risk. 

Filtering is another option, but, again, you are handling large amounts of 
formalin so risk high exposure and possible spills. We tried it and it was too 
labor intensive, the filter was constantly clogging and we had to keep large 
carboys around that seem to just leak due to the poor quality of the spigots. 
Again, we abandoned  it. 

The other questions are, who is going to do this (s EHS volunteering their 
staff?!?!)? Is there space to do it (waste carboys, plus filled carboys, plus 
transporting)? I can bet your lab does not have space or personnel for this 
procedure. Does EHS? 

Tim Morken
Pathology Site Manager, Parnassus
Supervisor, Electron Microscopy/Neuromuscular Special Studies Department of 
Pathology UC San Francisco Medical Center


-Original Message-
From: Martha Ward-Pathology via Histonet 
[mailto:histonet@lists.utsouthwestern.edu] 
Sent: Wednesday, December 20, 2017 11:00 AM
To: histonet@lists.utsouthwestern.edu
Cc: histo...@lists2.utsouthwestern.edu
Subject: [Histonet] Recycling formalin

I am posting this for our Histology lab manager:

Is anyone recycling formalin and if so, exactly what are you using the recycled 
formalin for?Putting it on processors, adding to specimens, etc?
Apparently our EHS department is pushing this.


Also thank everyone for their help with our questions about unstained slides 
for prostate biopsies.It was very helpful.

Thanks in advance for your help with our recycled formalin question.

Martha Ward
Wake Forest Baptist Health
336-716-2109
___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet


___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet


[Histonet] Recycling formalin

2017-12-20 Thread Martha Ward-Pathology via Histonet
I am posting this for our Histology lab manager:

Is anyone recycling formalin and if so, exactly what are you using the recycled 
formalin for?Putting it on processors, adding to specimens, etc?
Apparently our EHS department is pushing this.


Also thank everyone for their help with our questions about unstained slides 
for prostate biopsies.It was very helpful.

Thanks in advance for your help with our recycled formalin question.

Martha Ward
Wake Forest Baptist Health
336-716-2109
___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet


[Histonet] Hard Water and Auto Stainers

2017-12-20 Thread Sandra Cheasty via Histonet
Hello all,
Thanks for everyone's response to error 90 on our Prisma 
stainer; it has been cleared! Does anyone use filtered, softened, or DI water 
in their H stainer to avoid clogging up the system?
Thanks!
Sandy

Sandra J. Cheasty, HT (ASCP)
Histology & Necropsy Supervisor
UW-Madison, School of Veterinary Medicine

___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet