RE: [Histonet] Processor question - alcoholic formalin vs. buffered formalin

2013-04-30 Thread Hannen, Valerie
  We use 10 %  Neutral Buffered Formalin. It is a wonderful fixative and gives 
us really great fixation.
It does sound like you are overly dehydrating your tissues.

Valerie A. Hannen, MLT(ASCP),HTL,SU(FL)
Histology Section Chief
Parrish Medical Center
951 N. Washington Ave.
Titusville, Florida 32976
Phone:(321) 268-6333 ext. 7506
Fax: (321) 268-6149
valerie.han...@parrishmed.com



-Original Message-
From: histonet-boun...@lists.utsouthwestern.edu 
[mailto:histonet-boun...@lists.utsouthwestern.edu] On Behalf Of Lake, Kim S
Sent: Tuesday, April 30, 2013 11:31 AM
To: histonet@lists.utsouthwestern.edu
Subject: [Histonet] Processor question - alcoholic formalin vs. buffered 
formalin

I attended the Tri-State Symposium in Dubuque, IA last week (which was 
fantastic) and sat in on a Boot Camp for Histology workshop.  At this 
workshop I realized that some problems we have been having with sectioning 
could be due to our processor set up.  

Before microtomy we have to soak our blocks for 10-15 minutes, and I suspect 
that this is because we are over dehydrating our tissues in the during 
processing.  Our first solution, which the cassettes sit in for about 8 hours 
until the overnight processing begins, is alcoholic formalin, which is 450ml 
buffered formalin, 900ml water, and 3300ml 95% alcohol.  Our processor is a 
Leica TP 1050 that has been chugging along for about 20 years now.

The reason we use alcoholic formalin instead of just normal buffered formalin 
has been lost to the mists of time, and I was wondering what you all use as a 
holding solution on your processor, and why.

Thanks!

Kim Lake
Laboratory Manager
Oral Pathology Laboratory
University of Iowa College of Dentistry
Phone (319) 384 4433
Fax (319) 353 5569

___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet

=This email is intended solely for the use 
of the individual to
whom it is addressed and may contain information that is
privileged, confidential or otherwise exempt from disclosure
under applicable law. If the reader of this email is not the
intended recipient or the employee or agent responsible for
delivering the message to the intended recipient, you are
hereby notified that any dissemination, distribution, or
copying of this communication is strictly prohibited. If you
have received this communication in error, please immediately
delete this message. Thank you
=
___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet


RE: [Histonet] Processor Question

2012-02-21 Thread Davide Costanzo
Design flaw in the screen display. It is in the way of the chamber when
opening chamber. If your not careful you will break the screen. Happens
fairly often.

Sent from my Windows Phone
From: Gauch, Vicki
Sent: 2/21/2012 9:18 AM
To: histonet@lists.utsouthwestern.edu
Subject: [Histonet] Processor Question
Hi everyone,
We are in the market for new processors...and I was wondering if
anyone could give me some pros and cons for the Tissue Tek VIP 6
tissue processor - how reliable are they? Ease of use ? Any known
problems?  Tissues process well?   You knowall the usual questions
we all ask for new equipment.

Thanks in advance for your help,

Vicki Gauch
AMCH
Albany, NY



___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet

___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet