Re: [ccp4bb] def.site file of HKL2000 for Bruker X8 proteum

2008-02-19 Thread Bram Schierbeek
Dear Shi, The def.site file is specific for each individual CCD detector, so you cannot use any other def.site file from any other user of an X8Proteum system. If you do, indexing and hence processing will not work as it should. If you have a HKL2000 license, there should be a def.site present

[ccp4bb] def.site file of HKL2000 for Bruker X8 proteum

2008-02-19 Thread shi jiahai