Re: [PSES] IEC 61000-4-4 test setup for EUT which have casters

2013-05-24 Thread John Woodgate
In message <20130524.231204.219787274.vef00...@nifty.ne.jp>, dated Fri, 24 May 2013, T. Sato writes: Maybe a slightly silly question, but I want to hear your opinions. IEC 61000-4-4:2012 clause 7.3.1 says: Floor standing EUTs ... shall be placed on a ground reference plane and shall be ins

Re: [PSES] IEC 61000-4-4 test setup for EUT which have casters

2013-05-24 Thread Kunde, Brian
My first choice would be to test floor standing products sitting 0.1 m above the ground reference plane (GRP) on a support and don't consider the casters. However, some floor standing equipment is very heavy and can be difficult if not dangerous to try and pick up to set it on a 0.1m support. In

[PSES] IEC 61000-4-4 test setup for EUT which have casters

2013-05-24 Thread T.Sato
Colleagues, Maybe a slightly silly question, but I want to hear your opinions. IEC 61000-4-4:2012 clause 7.3.1 says: Floor standing EUTs ... shall be placed on a ground reference plane and shall be insulated from it by an insulating support with a thickness of (0,1 +/- 0,05) m including no