Re: [Histonet] Blank Spots on IHC- Bubbles on Bond

2020-08-14 Thread Morken, Timothy via Histonet
Greg, thanks! I have a leica engineer coming in to day to look at it, specifically the fluidics system, so hopefully he’ll see what you mention and fix it. Tim Morken Supervisor, Electron Microscopy/Neuromuscular Special Studies Department of Pathology UC San Francisco Medical Center From:

Re: [Histonet] Blank Spots on IHC- Bubbles on Bond

2020-08-13 Thread Greg Dobbin via Histonet
Hi Tim, The usual source of bubbles is a deteriorating syringe or lines or valve. Have you observed the syringe while running the Clean Fluidics maintenance function? There can be some "micro" bubbles in the syringe on first draw of a reagent however, if working properly these tiny bubbles quickly