Greg, thanks!
I have a leica engineer coming in to day to look at it, specifically the
fluidics system, so hopefully he’ll see what you mention and fix it.
Tim Morken
Supervisor, Electron Microscopy/Neuromuscular Special Studies
Department of Pathology
UC San Francisco Medical Center
From:
Hi Tim,
The usual source of bubbles is a deteriorating syringe or lines or valve.
Have you observed the syringe while running the Clean Fluidics maintenance
function? There can be some "micro" bubbles in the syringe on first draw of
a reagent however, if working properly these tiny bubbles quickly