We are moving out Peloris tissue processors to another room, so we have to
validate. I was wondering what other labs do as a validation for this type of
move? Do you run tissue on every progam utilizing each processor for at least
one of the protocols? Any help would be appreciated.
thanks,
This seems like overkill. Just moving a processor previously validated should
not require re-validation, just verify the programming and make certain the
paraffins are melted completely.
This may not apply to all processors, but for all Sakura machines manufactured
since 1980,I would not
, Westmead