RE: [Histonet] Grossing Hood Velocity Requirements

2010-12-17 Thread Maria Katleba
[mailto:histonet-boun...@lists.utsouthwestern.edu] On Behalf Of Feher, Stephen Sent: Friday, December 17, 2010 1:05 PM To: thisis...@aol.com; histonet@lists.utsouthwestern.edu Subject: RE: [Histonet] Grossing Hood Velocity Requirements We purchased Thermo's Gross Lab Senior grossing stations. A ru

RE: [Histonet] Grossing Hood Velocity Requirements

2010-12-17 Thread Feher, Stephen
We purchased Thermo's Gross Lab Senior grossing stations. A rule of thumb is for these to pull at 450-500 CFM. If you purchase the hoods that include the fan motor, these are factory set to this. We hooked ours in to the hospital ventilation system so it was a bit trickier to achieve this veloci