You've gotten gr
You've gotten great advice so far, but if it doesn't help--our problem was
that the specimen head was loose--even when it was locked, it would move
very slightly. This has to be fixed by servicing it, unfortunately. It was
caused by our MD/PhD student trying to adjust the specimen head when it wa