Re: [Ifeffit] Lattice parameters: EXAFS vs. XRD

2009-12-30 Thread Matt Newville
Hi Scott, I believe we had a conversation about this last January. XAFS is not sensitive to the crystallographic lattice constants. It measures the spacing between atoms. Because of thermal vibrations and other disorder terms, the average distance between atoms is larger than the distance

Re: [Ifeffit] Lattice parameters: EXAFS vs. XRD

2009-12-30 Thread Frenkel, Anatoly
A correction to Matt's email: In random alloys, smaller size atoms and larger size atoms are at different average distances (measured by EXAFS and other local-structure-sensitive techniques, e.g., XRD/PDF) that are, respectively, smaller and larger than the distance between the average lattice

Re: [Ifeffit] Lattice parameters: EXAFS vs. XRD

2009-12-30 Thread Scott Calvin
Thanks, Matt--you give a complete and satisfying discussion of this on January 23 on this list. I forgot about that because it came at the tail end of a long discussion as to whether C3 could ever be 0, but I suspect what you said then was rattling around in the back of my head and only