Hi Scott and Bruce,
In continuation with earlier email, here, I am just briefing the
observation on the analyzing systems. In my case, reference sample (bulk
YbVO4) itself takes large second so2 with huge uncertainty(2.04 +/- 1.36) to
get a satisfactory fit. Further, this parameter is found to cor
Hi Bruce,
I agree with your comments. Side phase with less than 5% in total
composition cannot be easily detected using regular XRD but comprehensively
using EXAFS. Here, i have also recorded bulk sample, similar to
nanoparticles composition, which is completely free from contamination.
>From thi