Eberline made the RM-14 and E-140 with only 3 scales to 50K CPM to avoid
this problem.  Friskers do not get a lot of use on the higher ranges.  At
SONGS, we ordered the 177 count rate with only 3 scales for the same reason.
You could retrofit to a 3 position switch, and scrape off the 1K indicator.
We also purchased about 40 Ludlum 2-1 friskers, with 3 scales and battery
only like the E-140.  The speaker is inside, and they are very easy to
calibrate and maintain (and they also bounce good and keep working). 

 

Roger Schanzenbach, CHP

 

 

 

From: Powernet [mailto:powernet-boun...@hpspowernet.org] On Behalf Of
Faulkenberry, Bill via Powernet
Sent: Tuesday, May 1, 2018 5:28 AM
To: 'powernet@hpspowernet.org' <powernet@hpspowernet.org>
Cc: Faulkenberry, Bill <bill.faulkenbe...@duke-energy.com>; Droppers, Amanda
Jo <amanda.dropp...@duke-energy.com>; Nicholson, Harry B
<harry.nichol...@duke-energy.com>; Steinmeyer, Tony
<tony.steinme...@duke-energy.com>; Wehrman, Libby
<libby.wehr...@duke-energy.com>; Blanton, John D
<john.blan...@duke-energy.com>; Sullivan, Danny
<danny.sulli...@duke-energy.com>; Barnhill, Carlette T
<carlette.barnh...@duke-energy.com>; Stevenson, Amanda Padgett
<amanda.steven...@duke-energy.com>; Warren, Eric M
<eric.war...@duke-energy.com>; Jackson, Clyde
<clyde.jack...@duke-energy.com>; Brantley, Cindy
<cindy.brant...@duke-energy.com>; Belfer, Brad
<brad.bel...@duke-energy.com>; Branham, Christy Munn
<christy.bran...@duke-energy.com>; James, Russell F
<russell.ja...@duke-energy.com>; Vincelli, John C
<john.vince...@duke-energy.com>; Galloway Jr., Gary Ross
<gary.gallow...@duke-energy.com>; Cerame, William P
<william.cer...@duke-energy.com>; Cuthbertson, Henry D
<henry.cuthbert...@duke-energy.com>; Smith, Joe W
<joe.sm...@duke-energy.com>; Parker, Michael Leon
<michael.park...@duke-energy.com>; Davis, Ken <ken.da...@duke-energy.com>;
Prosser, Terry Wyman <terry.pros...@duke-energy.com>; Robinson, Lagrone
Dexter <lagrone.robin...@duke-energy.com>; Huddleston, Laura Beth
<laura.huddles...@duke-energy.com>; Norton, Miles Rukean
<miles.nor...@duke-energy.com>; Chandler, William Zachary
<william.chand...@duke-energy.com>; Ceva, Celeste Marie
<celeste.c...@duke-energy.com>; Mc Call, Timothy A
<timothy.mcc...@duke-energy.com>
Subject: [Powernet] Ludlum Model 177 w/44-9 probe dead time

 

Duke Energy uses the Ludlum model 177 count rate meter with the Ludlum model
44-9 probe. Duke Energy recently learned that the Ludlum model 44-9 probe
has a manufacturer stated dead time of 80 microseconds. This would yield a
theoretical count rate loss of 67% at 500,000 cpm (the top of the X1K
scale). Actual dead time measurements made by Duke Energy yielded slightly
better results of 56% count rate loss at 500,000 cpm.  Duke Energy would
like to know the following:

 

1)    Does your site or utility use the Ludlum model 177 count rate with the
model 44-9 probe?

 

2)   Do you have any kind of restriction on the X1K scale due to probe dead
time? If so, what?

 

3) Do you perform any kind of dead time correction for the X1K scale during
calibration? If so, what?

 

4) Do you perform any kind of dead time correction for the X1K scale for
field measurements? If so, what?

 

Thanks,

Bill Faulkenberry, NRRPT

Supervising Scientist, EnRad Labs

Duke Energy

 



 

13339 Hagers Ferry Rd

Mail Stop MG03A2

Huntersville, NC  28078

Office: 980-875-6893

Mobile: 803-627-4832

 

_______________________________________________
Powernet mailing list
Powernet@hpspowernet.org
http://hpspowernet.org/mailman/listinfo/powernet_hpspowernet.org

Reply via email to