I was wondering with so many 10 mhz sources running.or even a single cesium
and gps receiver running, as I have now..and the 10 mhz is also routed to a
clock driver system (leitch).I cannot receive 10mhz wwv broadcasts at all
anytime,I am using a old radio shack sw radio on batt power..I tried
-nuts] GPS Locked and Unlocked Performance Comparison
David Welch wrote:
I was wondering with so many 10 mhz sources running.or even a single
cesium
and gps receiver running, as I have now..and the 10 mhz is also routed
to a
clock driver system (leitch).I cannot receive 10mhz wwv broadcasts
on 2.5, 5 and 15 MHz. 15 MHz is good
during
mid day and 5 MHz is quite good at night time. 2.5 MHz is only good at
night and
a little tough to copy unless you have a good antenna or are living quite
close
to it.
BillWB6BNQ
David Welch wrote:
I was wondering with so many 10 mhz
- Original Message -
From: David Welch [EMAIL PROTECTED]
To: Dave Carlson [EMAIL PROTECTED]
Sent: Sunday, February 10, 2008 9:24 PM
Subject: Re: [time-nuts] cs beam tube failure???5071a tube is dead now...
Dave..
I did get a case # with agilent and they said to call symetricom
___
- Original Message -
From: David Welch [EMAIL PROTECTED]
To: Discussion of precise time and frequency measurement
time-nuts@febo.com
Sent: Friday, February 08, 2008 9:59 AM
Subject: [time-nuts] cs beam tube failure???
Dave,
I had a similar
Dave,
I had a similar concern about a brand new 5071a I just purchased,This unit
is dated may of 05 and was first powered up Jan 2 of this year..the ion pump
has remained constant at 6 ua but I the electron mult volt has been dropping
about 1-2 volts per day since then.it started out at 1525 and
Dave,
I had a similar concern about a brand new 5071a I just purchased,This unit
is dated may of 05 and was first powered up Jan 2 of this year..the ion pump
has remained constant at 6 ua but I the electron mult volt has been dropping
about 1-2 volts per day since then.it started out at 1525 and
Does anyone know how much a decent 5071a goes for these days..I hear the tube
new is about 15k..would 27K be a good price for a new unit?
I think upgrading from 5061b to 5071a might be better investment due to the high
cost of a new tube..
___
time-nuts
Just dug uot my 5061a out of storage and left it on onernight..I have never
checked this unit after purchasing it 4 yrs ago.
when I first powered up the Cs oven was pegged at 50 and has slowly dropped to
10..but the cs oven is not activing yet ..does anyone know how low it has to
get before it
- Original Message -
Hello.I was wondering if any of you guy know a quick easy way to eval tube
condition on a 5071a ,these are several years of disuse on these units and I
was thinkng of purchsahing a few .does time to lock have anything to do with
tube health or are there other
I am new to this forum and was wondering if someone would know how to check if
the tube in a agilent 5071a is any good.these units are new in box but have
been in storage many years(bad idea)..is there anything to look at during start
up to give you an indication of outgassing ar vacum loss?
11 matches