Re: [time-nuts] Restoration of a 5065A - Continued (Ulf Kylenfall)

2017-09-12 Thread Scott McGrath
Hi Ulf Also you noted the bulb had the thermionic device, That's used for remigrating the Rb into the reservoir there is a procedure for doing it in the manuals on the web. It can take up to a week to draw all the Rb back into the reservoir after long inactivity.Just make sure you use a

Re: [time-nuts] Restoration of a 5065A - Continued (Ulf Kylenfall)

2017-09-12 Thread paul swed
Totally agree with Scott. On the lamp thats good it lights. If you search older threads on Time-nuts you will see that on the FRS rubidiums you can heat the lamp with a heat gun and re-migrate the RB off of the glass. It makes them re-usable and extends the life very nicely. I do not suggest doing

Re: [time-nuts] Restoration of a 5065A - Continued (Ulf Kylenfall)

2017-09-12 Thread Scott McGrath
I would say fix both units The color of the bulb is simply Rb vapor condensed on the walls of the envelope. 10811 units are fairly common at good hamfests and ebay Content by Scott Typos by Siri > On Sep 12, 2017, at 4:54 PM, Ulf Kylenfall via time-nuts > wrote: > > > What does the communit

[time-nuts] Restoration of a 5065A - Continued (Ulf Kylenfall)

2017-09-12 Thread Ulf Kylenfall via time-nuts
Picked up a scrapped 5065A tonight. It was an early unit but a later one than the one I amtrying to fix. The XO was gone. From what I could see, it had beena 10811 type with adapterĀ  since there was no 115V wiring. The RV Cavity was loose and I think that what was in this unit mayhave been a rep