-----BEGIN PGP SIGNED MESSAGE-----
Hash: SHA1

Hello Sam--

> 
> I am trying to get the distance restraints from ROESY measurement for
> xplor-nih, but I have question about how the calibration (binning) of the
> distances should be carried out. Do I follow the same calibration techniques 
> as
> in NOESY? 
> 
>  

Here is Marius' advice:

+ No reason, but: ROEs are NOT affected by spin-diffusion, so the intensities
+ should be directly proportional to r-6.
+ So if one wanted to, one could be a lot more restraining with ROEs.

Charles
-----BEGIN PGP SIGNATURE-----
Version: GnuPG v1.4.9 (GNU/Linux)
Comment: Processed by Mailcrypt 3.5.8+ <http://mailcrypt.sourceforge.net/>

iEYEARECAAYFAkpWRnIACgkQPK2zrJwS/lZYzwCeOwh9YYhSqchT1ydZmOdUz+H0
duwAn1p9qPsVEYmufbzDyG0hZCHjWFrb
=bEZY
-----END PGP SIGNATURE-----

Reply via email to