On Thu, 2011-10-27 at 15:36 +0200, George M. Sheldrick wrote: > In non-continuous mode, the goniometer has > to accelerate at the start of a frame and decellerate at the end, then > wait for > the frame to be read.
Someone should be able to confirm this, but I was under impression that at synchrotrons the acceleration/decelaration occurs outside phi_start-phi_end range to allow for constant speed when the shutter opens. -- "I'd jump in myself, if I weren't so good at whistling." Julian, King of Lemurs