I'd hang a 1N914 diode across the output of the Siglent generator and tune it to 25 MHz.  The 2nd harmonic should be strong enough.

73

Bob, K4TAX


On 5/2/2019 5:26 PM, Frank O'Donnell wrote:
Which of the following two signal sources would enable the best results from the KX3's extended temperature calibration?

--Elecraft's XG50 board, which puts out 49.380 MHz ± 5 ppm.

-- Siglent SDG 2042X waveform generator, with input from a Trimble Thunderbolt GPSDO. Maximum frequency on the Siglent is 40 MHz.

The calibration instructions call for using a signal source close to 50 MHz. Is 40 MHz close enough? The XG50 is much closer to 50 MHz, whereas the Siglent/Trimble potentially has the capability of being considerably more accurate. I'm interested in squeezing as much frequency stability out of the KX3 as possible.

Thanks and 73,

Frank O'Donnell K6FOD