Hi,
 
I am going to be in the same boat soon, especially regarding obtaining 
instrument parameters for a new instrument.  But one thing I do know, is that 
to get the polarization parameter, quartz is not a great standard.  For 
polarization factors, you need something really simple, with heavy atoms and 
all atoms on centers of symmetry, and with thermal parameters very well 
constrained, so that you can REFINE the polarization parameters for your 
machine, particularly when incindent beam monochromators are used.
 
The only problem is, I can't remember what the favorite standard is, or where 
it can be purchased.  I will need to know myself, so I will ask, on behalf of 
Blaise and myself as well, what is the best standard for refining the 
polarization of an instrument with monochromatized beam? Thank you!
 
- Kurt

________________________________

From: Mibeck, Blaise [mailto:[EMAIL PROTECTED]
Sent: Wed 9/17/2008 12:22 PM
To: rietveld_l@ill.fr
Subject: Beginer problems, difficulty charecterizing our diffractometer...



 

I am re-learning GSAS to bring Rietveld to my department (for the first time). 
We have an old Phillips Xpert. 

 

I am trying to refine a quartz standard to acquire my profile and instrument 
parameters for this instrument and have yet to get my Chi^2 below 500. The 
instrument is not in my direct control. I have caught and asked them to correct 
a few problems already (dwell time too low, aluminum sneaking into the beam, 
etc...) and this has helped. I would like not to annoy them any more than I 
have too. 

 

At this point I think my high Chi^2s are mainly due to low angle asymmetric 
peaks that my fit is not able to copy - the asym is not terrible but may be 
messing me up. I am getting better results with peak profile functions 3 and 4, 
but unable to get Chi^2 below 500. My scans go up to 70 degrees 2theta.

 

I would like to tell the people caring for the instrument that my problem is on 
their end, but I am not confident in my own refining skills to say this. When I 
work with the tutorials or standard data from my graduate school experiment I 
am ok - I think I am proceeding in a reasonable way, although I can get stuck 
here and there. 

 

I would like to find out if I am doing something incorrectly or if the problem 
is instrument related. I hate to bug one of you but wonder if I could get 
someone to look at my work?? Is there a better forum for asking for this kind 
of help? 

 

Kindest regards to all of you,

Blaise

 

 

 

 

 

 

* * * * * * * * * * * * * 

Blaise Mibeck 

Research Scientist

Energy & Environmental Research Center

University of North Dakota

15 North 23rd Street, Stop 9018

Grand Forks, ND 58202-9018

 

Phone: (701) 777-5077

Email: [EMAIL PROTECTED]

 

 

 