http://www.kingtigertkdwilson.com/modules/mod_wdbanners/myfriends.php?central164.gif
___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet
Hello:Are any of the Meditech users having the pathologists order their levels
in the computer?We would like to implement this procedure. Any suggestions.
:)ThanksSheila
___
Histonet mailing list
Is there anyone out there that does repairs on Sakura equipment and works
for themselves
in the Texas area.
___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
http://lists.utsouthwestern.edu/mailman/listinfo/histonet
I don't know about implants but we were having a terrible time with bone
marrows coming off. Usually they came off during the HEIR using the steamer.
At that point we were placing our slides lying flat in a 55-60ยบ oven overnight
which usually helps tremendously with difficult tissue. This was
Sheila asks:
Hello: Are any of the Meditech users having the pathologists order their
levels in the computer? We would like to implement this procedure. Any
suggestions?
Never seen this - but it sounds like a nifty little time-waster for
all concerned.
I usually examine the paraffin block
Region 1 Symposium Announcement
The New York State Histotechnological Society is hosting the 2012 Region 1
Histotechnologist Symposium to be held at the Marriott Islandia Long Island on
April 27th and 28th 2012.
The program/registration, exhibitor and hotel/travel information is available
This is an age old problem, bone comes off easily especially during HIER,
which is why I have adapted all my bone IHC abs to use EIER instead of HIER,
usually with pepsin or proteinase K. Also, it helps to heat the slides flat
on a slide warmer at 45dc overnight. If all else fails I use elmer
Original Message --
Message: 9
Date: Tue, 14 Feb 2012 16:31:29 -0600
From: Vickroy, Jim vickroy@mhsil.com
Subject: [Histonet] New DNA Probes for Kappa and Lambda
To: histonet@lists.utsouthwestern.edu
histonet@lists.utsouthwestern.edu
Message-ID:
Can anyone tell me where I could find a small (6-7 inches long) alcohol
hydrometer? Needs to read to 200% proof.
All I'm finding is a 12 size and I would like to be able to test in a
100ml cylinder.
Thanks for your help,
Kim
Kimberly D. Kolman, HT, (ASCP)
Diagnostics 115
VA
I've never seen one that small.
Sent from the iPhone of Kim Tournear
On Feb 15, 2012, at 2:26 PM, Kolman, Kimberly D. kim.kol...@va.gov wrote:
Can anyone tell me where I could find a small (6-7 inches long) alcohol
hydrometer? Needs to read to 200% proof.
All I'm finding is a 12 size
Hello, Esteemed Histonetters!
I have a collaborator who asked me to drop his (mouse) bones into RNAlater for
RNA isolation, later. These animals were double labeled for histomorphometry as
well. He now wants to know if he can back them out of RNAlater and use them for
any histology-related
Hi all,
I am staining cryo sections of aorta for lipids with Oil Red O and
mounting my slides with Mowiol.
My question is how long will they last without fading and should they be
kept in a folder in the fridge?
Thanks for your thoughts,
Daniela
Daniela Bodemer
Research
Sometimes while cutting 5 micron paraffin sections...
I turn the wheel once and a nice section gets cut and lays down on my blade.
When I turn the wheel again, as the knife holder is traveling Upwards, the top
of my paraffin block will grab the ribbon and lift it off of the microtome
blade,
Checking the troubleshooting manual for the Leica 2525...
Problem: Specimen is picked up in the return stroke of the specimen arm.
Possible Cause: Static electricity charge may be built up on the knife holder
or specimen head.
Corrective Action: Clean components of microtome with alcohol
...Seems like you should give this a try
http://audacesidea.com.br/day.link.php?ykegoogleId=88or8
___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
From: del phillips dphill...@vetmed.lsu.edu
Subject: [Histonet] IHC and bone implants
Does anyone have any suggestions for staining bone implants? I am having
trouble with wash offs. I use charged slides and they are still washing off.
Thanks
Del
Del,