Hi All,
Does anyone know where I can purchase some spare vip3000 reagent bottles?
Regards
Mitchell Wan
0418 745 750
P.O. BOX 2200,
Runcorn,
Brisbane,
QLD 4113
___
Histonet mailing list
Histonet@lists.utsouthwestern.edu
Hi Mike,
Edit the 'dummy' run and make station 3 with '0' time.
It is time based and has a safety mechanism. That's why it pumps the previous
station. Then test again.
A couple things may be an issue and need to be tested.
1. A hole, leak in the line of station 3.
2. Volume in reservoir bottle