RE: [Histonet] VIP 2000

2010-07-09 Thread Bell, Lynne
Error code 73: Symptom: Rotary valve does not stop at the required station after 2 times rotation Probable cause: 1. Photointerrupter defective 2. Positioning disc broken 3. Photointerrupter for position detection disturbed by strong lightsource

Re: [Histonet] VIP 2000

2010-07-09 Thread Drew Meyer
You can always call Sakura... they'll at least tell you what the codes means... Drew On Fri, Jul 9, 2010 at 10:15, wrote: > > Hello all, > > I recently purchased a refurbished VIP. ; This would have been the > 3rd time I have run it, but wait...diagnose flas hes and I get an > er

[Histonet] VIP 2000

2010-07-09 Thread sgoebel
Hello all, I recently purchased a refurbished VIP. = ; This would have been the 3rd time I have run it, but wait...diagnose flas= hes and I get an error code 73. Of course the manual doesn't say what= this is or how to fix it, and I can't seem to find anything online eithe