Error code 73:
Symptom: Rotary valve does not stop at the required station after 2 times
rotation
Probable cause: 1. Photointerrupter defective
2. Positioning disc broken
3. Photointerrupter for position detection disturbed by
strong lightsource
You can always call Sakura... they'll at least tell you what the codes
means...
Drew
On Fri, Jul 9, 2010 at 10:15, wrote:
>
> Hello all,
>
> I recently purchased a refurbished VIP. ; This would have been the
> 3rd time I have run it, but wait...diagnose flas hes and I get an
> er
Hello all,
I recently purchased a refurbished VIP. = ; This would have been the
3rd time I have run it, but wait...diagnose flas= hes and I get an
error code 73. Of course the manual doesn't say what= this is or how
to fix it, and I can't seem to find anything online eithe