Dear all,
I found the solution analysing headers of TEST data. It was an stupid
mistake in AUTOPIC.INP file. An "H" was mising in the station description!!
if channel name is HHZ, HHN or HHE it must be:
*E001 HH Z 3 component
*E002 HH Z 3 component
*E005 HH Z 3 component
*E006 HH Z 3 component
Instead of:
*E001 H Z 3 component
*E002 H Z 3 component
*E005 H Z 3 component
*E006 H Z 3 component
Perhaps, these parameters should be better explained in future versions
of Seisan manual...
Best regards,
Mario
El 25/01/12 08:52, Jens Havskov escribió:
mario
you certainly havtested all possibilities! the most likely errro is
with parameter file, it must srictly have the format given lik
eeverythin in correct column and no tabs. E.g. station ALJ below is
wrongly formatted.
The "preprocess.inf" and "AUTOPIC.INF must be some leftover for very
early versions, all info should be in manual
please send data file, parameter file and waveform file if you do not
find the problem
jens
On Tue, 24 Jan 2012, Mario Ruiz Fernandez wrote:
Dear all,
I'm trying to use AUTOPIC program by typing Z on eev (9.0.1
distribution),
but command does anything on my dataset. I don't understand why
because with
example dataset coming in TEST database it works perfectly well!
I updated AUTOPIC.INP introducing the list of stations to be analysed:
% STATIONS TO USE IN THE PROCESSING
% ----------------------------------------------------------------------
*ACBG B Z 3 component
*ACLR B Z 3 component
*ALHU H Z 3 component
*ALJ H Z 3 component
*ANER B Z 3 component
*ARAC B Z 3 component
*ASCB B Z 3 component
*E001 H Z 3 component
*E002 H Z 3 component
*E005 H Z 3 component
*E006 H Z 3 component
*E007 H Z 3 component
*E008 H Z 3 component
*E009 H Z 3 component
*E010 H Z 3 component
*E011 H Z 3 component
*E012 H Z 3 component
*E013 H Z 3 component
*E015 H Z 3 component
*E017 H Z 3 component
*E018 H Z 3 component
*E019 H Z 3 component
etc...
But when I press Z, I only obtain the following message:
# 18 6 Nov 2008 12:50 45 L 37.019 -5.409 21.0 N .82 3.2bIGN
29 ? Z
*******************************************************
**** ****
**** AUTOPIC routine ****
**** ****
*******************************************************
/home/mario/Seisan_9.0.1/Seismo/REA/IBER_/2008/11/06-1250-45L.S200811
?? ?
waveform
file:/home/mario/Seisan_9.0.1/Seismo/WAV/IBER_/2008/11/2008-11-06-1250-45S.mseed
multiplexed
/home/mario/Seisan_9.0.1/Seismo/DAT/AUTOPIC.INP
Output written to autopic.out
Nothing else! Any "detasso.*" file is generated and "autopic.out" only
contains these 3 lines:
Reading waveforms from
file :/home/mario/Seisan_9.0.1/Seismo/WAV/IBER_/2008/11/2008-11-06-1250-45S.msee
d
Parameter settings in file
:/home/mario/Seisan_9.0.1/Seismo/DAT/AUTOPIC.INP
Parameter explanation in file:AUTOPIC.INF
I also did two tests to rule out any problem with our waveform data
and/or
with the file size (multiplexed mseed files containing 95 stations
with 3
channels each one with sampling rates of 100, 50 and 20sps):
1) I converted one of the mseed files to seisan format with
"wavetool" and
introduced the new waveform file in S-file. 2) I split the same mseed
file
in all thier channels and stations using "seisei" command and introduced
only the Z channel of best station in s-file, but the result is the
same.
Any pick is obtained.
I'm wondering if I'm doing any mistake on station description in
AUTOPIC.INP
file. Must they be sorted? Are spaces relevant? Where could I find
"preprocess.inf" and "AUTOPIC.INF" files to have a complete
explanation of
all the formats and parameters needed? I'm not able to find them in my
Seisan distributions (9.0.1, 8.3 or 8.2.1) ...
Best regards,
Mario
--
------------------------------------------------------------
Mario Ruiz Fernandez
Dep. d'Estructura i Dinamica de la Terra
Institut de Ciencies de la Terra 'Jaume Almera' ICTJA-CSIC
c/ Lluis Sole Sabaris s/n 08028 Barcelona (Spain)
Tel.: +34 93 409 54 10
Fax: +34 93 411 00 12
E-mail: [email protected]
www: http://www.ictja.csic.es/edt/
http://www.tdx.cat/handle/10803/1923
------------------------------------------------------------
--
------------------------------------------------------------
Mario Ruiz Fernandez
Dep. d'Estructura i Dinamica de la Terra
Institut de Ciencies de la Terra 'Jaume Almera' ICTJA-CSIC
c/ Lluis Sole Sabaris s/n 08028 Barcelona (Spain)
Tel.: +34 93 409 54 10
Fax: +34 93 411 00 12
E-mail: [email protected]
www: http://www.ictja.csic.es/edt/