In message <[EMAIL PROTECTED]>, "Ulrich Bangert" writes:

>But what surprises us completely is the fact that different physics
>packages need DIFFERENT thumbwheel settings to generate the SAME time
>scale as seen with the two devices available.

This is because rubidium vapour standards are not primary standards.

They don't have a "correct" frequency, only a very stable frequency.

The actual resonance frequency of a rubidium standard depends amongst
other things on the partial pressures inside the physics package.

-- 
Poul-Henning Kamp       | UNIX since Zilog Zeus 3.20
[EMAIL PROTECTED]         | TCP/IP since RFC 956
FreeBSD committer       | BSD since 4.3-tahoe    
Never attribute to malice what can adequately be explained by incompetence.