> Comparable to the Datron 4911.
The 4410 is based on ovenised LM329 reference = LM399 without a heater. You can expect drift by factor of 10 worse than Datron 4910/4911 or a good Fluke 732A. _______________________________________________ volt-nuts mailing list -- volt-nuts@febo.com To unsubscribe, go to https://www.febo.com/cgi-bin/mailman/listinfo/volt-nuts and follow the instructions there.