Hi Almu,

I have yet to see a data set with clear reflections that cannot be indexed with 
XDS. The most often reason for failure is some wrong input, most often 
ORGX/ORGY and ROTATION_AXIS, but of course OSCILLATION_RANGE, X-RAY_WAVELENGTH, 
DETECTOR_DISTANCE must also be correct.
If you cannot solve the problem, then I offer my help - make the raw data 
available to me, using some cloud service, and I'll take a look. I'll keep your 
data confidential of course.

best,
Kay

On Fri, 29 Nov 2019 16:06:53 +0100, Almudena Ponce Salvatierra 
<maps.fa...@gmail.com> wrote:

>Thank you Tim,
>
>I will try and let you know how it goes.
>
>All the best,
>
>Almudena
>
>El vie., 29 nov. 2019 a las 14:43, Tim Gruene (<tim.gru...@univie.ac.at>)
>escribió:
>
>> Dear Almudena,
>>
>> SPOT_MAXIMUM-CENTROID is probably the keyword in XDS.INP that you want to
>> increase in your situation. The default is 3. Try e.g. 5, and increase
>> slowly
>> to make sure that the suggested solution makes sense
>>
>> Best regards,
>> Tim
>>
>> On Friday, November 29, 2019 12:48:11 PM CET Almudena Ponce Salvatierra
>> wrote:
>> > Dear all,
>> >
>> > I have some data sets that don't want to be processed :p
>> >
>> > In one of them, when I look at IDXREF.LP I see virtually none of the
>> found
>> > spots were indexed and the reason is that they are "too far from the
>> > expected position". The spots are smeary and elongated, so not the
>> > prettiest.
>> >
>> > I have managed to process so far only one data set with decent statistics
>> > from another crystal harvested from the same drop, where the diffraction
>> > spots look better.
>> >
>> > I am trying to find in the xds wiki the keyword I should fine tune in
>> order
>> > to make those spots indexable.
>> >
>> > Could you help me please?
>> >
>> > Thank you very much in advance.
>> >
>> > Best wishes,
>> >
>> > Almu
>> >
>> > ########################################################################
>> >
>> > To unsubscribe from the CCP4BB list, click the following link:
>> > https://www.jiscmail.ac.uk/cgi-bin/webadmin?SUBED1=CCP4BB&A=1
>>
>> --
>> --
>> Tim Gruene
>> Head of the Centre for X-ray Structure Analysis
>> Faculty of Chemistry
>> University of Vienna
>>
>> Phone: +43-1-4277-70202
>>
>> GPG Key ID = A46BEE1A
>>
>
>########################################################################
>
>To unsubscribe from the CCP4BB list, click the following link:
>https://www.jiscmail.ac.uk/cgi-bin/webadmin?SUBED1=CCP4BB&A=1
>

########################################################################

To unsubscribe from the CCP4BB list, click the following link:
https://www.jiscmail.ac.uk/cgi-bin/webadmin?SUBED1=CCP4BB&A=1

Reply via email to