Dear Doo Nam Kim,

On 09.10.20 08:43, Doo Nam Kim wrote:
Since I don't know correct value of OSCILLATION_RANGE, I just changed to the 
example value (0.1) from (0.899999976, I know it is not a positive multiple of 
0.0001, but worked for ketone data).

without knowing the correct value for the OSCILLATION_RANGE, data processing might run through but produces non-sense results, independent of the data processing program you use. So, it is absolutely important to know the correct value for the oscillation range per frame.

Cheers,

Dirk.

--

*******************************************************
Dirk Kostrewa
Gene Center Munich
Department of Biochemistry, AG Hopfner
Ludwig-Maximilians-Universität München
Feodor-Lynen-Str. 25
D-81377 Munich
Germany
Phone:  +49-89-2180-76845
Fax:    +49-89-2180-76998
E-mail: dirk.kostr...@lmu.de
WWW:    www.genzentrum.lmu.de
        strubio.userweb.mwn.de
*******************************************************

########################################################################

To unsubscribe from the CCP4BB list, click the following link:
https://www.jiscmail.ac.uk/cgi-bin/WA-JISC.exe?SUBED1=CCP4BB&A=1

This message was issued to members of www.jiscmail.ac.uk/CCP4BB, a mailing list 
hosted by www.jiscmail.ac.uk, terms & conditions are available at 
https://www.jiscmail.ac.uk/policyandsecurity/

Reply via email to