As for the purpose of operation, EMC chambers need no grounding.
The shielding property happens by conduction of the shield material the
chamber is made from.

But as soon as mains filters are mounted on the wall, one has to cope with
leakage currents
(blind current) as large as 6 amp or more. This is definitely lethal on the
touch so I recommend a
grounding of 4-6mm2 copper in addition to the ground lead in the mains
connection.

I suggest leading both grounds to the building ground, as that is an
equipotential point, made for safety.

But again, an ungrounded chamber performs the same as a grounded one.

Gert Gremme


-----Oorspronkelijk bericht-----
Van: Price, Andrew (Leonardo, UK)
[mailto:andrew.p.pr...@leonardocompany.com] 
Verzonden: dinsdag 18 juli 2017 16:55
Aan: EMC-PSTC@LISTSERV.IEEE.ORG
Onderwerp: Re: [PSES] Chamber grounding [General Use]

Hi All

Which is the best to use?
Ground each EMC chamber via an earth spike or back to the transformer of the
supply to the building?
Which is best practice?

Regards
Andy



         Andrew Price
             Land & Naval Defence Electronics Division
             Prinicpal Environmental Engineer (EMC)

             Leonardo MW Ltd
             Sigma House, Christopher Martin Rd, Basildon SS14 3EL, UK
             Tel  EMC LAB : +44 (0)1268 883308
             Mobile: +44 (0)7507 854888
 
andrew.p.pr...@leonardocompany.com<mailto:andrew.p.pr...@leonardocompany.com
>
             leonardocomapany.com
HELICOPTERS / AERONAUTICS / ELECTRONICS, DEFENCE AND SECURITY SYSTEMS /
SPACE

* Please consider the environment before printing this email.




Leonardo MW Ltd
Registered Office: Sigma House, Christopher Martin Road, Basildon, Essex
SS14 3EL A company registered in England & Wales.  Company no. 02426132
********************************************************************
This email and any attachments are confidential to the intended recipient
and may also be privileged. If you are not the intended recipient please
delete it from your system and notify the sender.
You should not copy it or use it for any purpose nor disclose or distribute
its contents to any other person.
********************************************************************

-
----------------------------------------------------------------
This message is from the IEEE Product Safety Engineering Society emc-pstc
discussion list. To post a message to the list, send your e-mail to
<emc-p...@ieee.org>

All emc-pstc postings are archived and searchable on the web at:
http://www.ieee-pses.org/emc-pstc.html

Attachments are not permitted but the IEEE PSES Online Communities site at
http://product-compliance.oc.ieee.org/ can be used for graphics (in
well-used formats), large files, etc.

Website:  http://www.ieee-pses.org/
Instructions:  http://www.ieee-pses.org/list.html (including how to
unsubscribe) List rules: http://www.ieee-pses.org/listrules.html

For help, send mail to the list administrators:
Scott Douglas <sdoug...@ieee.org>
Mike Cantwell <mcantw...@ieee.org>

For policy questions, send mail to:
Jim Bacher:  <j.bac...@ieee.org>
David Heald: <dhe...@gmail.com>

-
----------------------------------------------------------------
This message is from the IEEE Product Safety Engineering Society emc-pstc 
discussion list. To post a message to the list, send your e-mail to 
<emc-p...@ieee.org>

All emc-pstc postings are archived and searchable on the web at:
http://www.ieee-pses.org/emc-pstc.html

Attachments are not permitted but the IEEE PSES Online Communities site at 
http://product-compliance.oc.ieee.org/ can be used for graphics (in well-used 
formats), large files, etc.

Website:  http://www.ieee-pses.org/
Instructions:  http://www.ieee-pses.org/list.html (including how to unsubscribe)
List rules: http://www.ieee-pses.org/listrules.html

For help, send mail to the list administrators:
Scott Douglas <sdoug...@ieee.org>
Mike Cantwell <mcantw...@ieee.org>

For policy questions, send mail to:
Jim Bacher:  <j.bac...@ieee.org>
David Heald: <dhe...@gmail.com>

Reply via email to