This link
<https://files.ettus.com/manual/page_calibration.html#calibration_self_utils>
indicates that the antenna ports are to remain unconnected. Since there is
no internal loopback path, I believe that the calibration algorithm
utilizes the "leakage" signal which is apparently strong enough such that
it can be minimized with the appropriate calibration adjustments.
Rob


On Thu, Feb 29, 2024 at 11:16 AM zhou via USRP-users <
usrp-users@lists.ettus.com> wrote:

> Hi Cédric,
>
> It was instructed by Ettus engineer when we started to use USRPs a few
> years ago. This is because there is no internal loopback path in circuit in
> X310. By the way, we are using UBX daughterboard. Maybe this is not
> documented.
>
> Regards,
> Hongwei
>
>
>
> On Thursday, 29 February 2024 at 15:05:37 GMT, Cédric Hannotier via
> USRP-users <usrp-users@lists.ettus.com> wrote:
>
>
> Hi Hongwei,
>
> On 2024-02-26 10:06 +0000, zhou wrote:
> > For X310 USRPs, you need to loopback connect the antenna ports.
>
> Could you direct me to where you got that info?
> I never encountered it, and I am not able to find it in the Ettus docs.
> Doing proper calibration is important,
> so we should ensure that the Ettus docs give the correct procedure...
>
> Kind regards
>
> --
>
> Cédric Hannotier
> _______________________________________________
> USRP-users mailing list -- usrp-users@lists.ettus.com
> To unsubscribe send an email to usrp-users-le...@lists.ettus.com
> _______________________________________________
> USRP-users mailing list -- usrp-users@lists.ettus.com
> To unsubscribe send an email to usrp-users-le...@lists.ettus.com
>
_______________________________________________
USRP-users mailing list -- usrp-users@lists.ettus.com
To unsubscribe send an email to usrp-users-le...@lists.ettus.com

Reply via email to